4.7.4 Nurse Call Setup
Select NURSE CALL SETUP >> in USER MAINTAIN menu. The following menu appears.
Figure 4-26 Nurse Call Setup
You can perform the following settings:
SIGNAL DURATION
Two options are available: PULSE, and CONTINUUM.
1. PULSE
When pulse is selected, a nurse call signal is a pulse signal lasting 1s. When multiple alarms
occur simultaneously, only one pulse signal will be output. If an alarm comes out before the
previous alarm is cleared, another pulse signal will be output.
2.
CONTINUUM
When continuum is selected, the duration of a nurse call signal is the same with the alarm,
namely, from the time that the alarm occurs to the time it disappears.
SIGNAL TYPE
1.
NORMAL OPEN: Select this option when the hospital’s call system is set to NORMAL
OPEN.
2.
NORMAL CLOSE: Select this option when the hospital’s call system is set to
NORMAL CLOSE.
4-24