Shenzhen PM-9000 Express Manuel de l'opérateur Page 81

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 339
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 80
ALARM SOUND ENABLE: ALARM VOL can be set to 0.
OFF DISABLE: ALARM VOL cannot be set to 0.
MODE 1: the alarm sound interval of high level alarm is 8s and
AUDIO OFF
that of medium level alarm is 24s.
MODE 2: the alarm sound interval of high level alarm is 3s and
that of medium level alarm is 14s.
PARA ALM TYPE LATCH, UNLATCH
ECG LEAD OFF Alarm level: HIGH, MED, LOW.
SpO2 SENSOR OFF Alarm level: HIGH, MED, LOW.
WARNING
Select the setting of ALARM SOUND OFF with caution.
NOTE
In the case that alarm volume is set to 0, if you set ALARM SOUND OFF to
DISABLE, alarm volume change to the default.
4.7.2 IP Address Setup
When the monitor is connected with the central monitoring system, and the NET TYPE is
CMS+, you need to set the IP address of your monitor. Select IP ADDRESS SETUP in USER
MAINTAIN menu. The following menu appears. For details, please contact with the
technician responsible for the central monitoring system in your hospital.
Figure 4-24 IP Address Setup
4-22
Vue de la page 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 338 339

Commentaires sur ces manuels

Pas de commentaire