If you start measurement in the 0 - 30s level, it will be
switched to 0 - 60s level automatically if the measurement
cannot be finished within 30 seconds. After the switch, no
further adjustment can be made to the X-axis.
This option allows you to enter WINDOW FOR C.O. EDIT.
This option allows you exit WINDOW FOR CO
MEASUREMENT.
1.
The cardio output measurement can be started when the prompt information “Ready for
new measurement” appears in the window.
2.
Select the START option, and then start the injection.
3.
The thermodilution curve, the current blood temperature and the injectate temperature
are displayed in the process of the measurement.
4.
The curve drawing stops automatically when the measurement is completed. At the
same time, the C.O. (Cardiac output) and the C.I. (Cardiac Index) are calculated and
displayed in the window. The monitor also displays the CO value in the CO parameter
window.