Shenzhen PM-9000 Express Manuel de l'opérateur Page 240

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 339
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 239
17.2.4 CO2 User Maintain Menu
Select CO2 USER MAINTAIN >> in USER MAINTAIN menu. The following menu
appears.
Figure 17-5 CO2 User Maintain Menu
The following information is displayed in the menu above.
CO2 The currently measured CO2 content.
Unit: %.
In the field at the right of the CO2 value, you can select a standard gas containing a certain
amount of CO2 for calibration.
Options: 3%, 4%, 5%, 6% and 7%.
BAROMETRIC The currently measured barometric pressure.
Unit: mmHg.
SENSOR TEMP
The currently measured temperature around the infrared
sensor.
Unit: ..
CUR PUMP RATE The currently measured pumping rate.
Unit: ml/min.
SET PUMP RATE
The sampling flow rate set by the user.
17-9
Vue de la page 239
1 2 ... 235 236 237 238 239 240 241 242 243 244 245 ... 338 339

Commentaires sur ces manuels

Pas de commentaire