Weak
No No 1 The ECG signal is too
weak to be detected.
Select a lead with better ECG
signal output. Check the
placement of the electrode or
replace the leadwire.
C.2.3 RESP Module Alarm Messages
Alarm message A B L Cause Measure
RESP
DISTURBED
No No 3 The module circuit is
disturbed.
If the problem occurs
continuously, restart the
monitor. If it still exits,
contact our company for
repair.
RR EXCEED No No 1 The circuit is disturbed and
the measurement is
inaccurate.
C-3