NOTE
Set the injectate temperature source properly in the CO setup menu. The
ON/OFF state of the injectate temperature source, when measurement is
finished, affects the CO calculation. Thus, do not change the injectate
temperature source before the measurement is over.
Monitor
Injectate
Cardiac Output
Cable
Thermodilution
Catheter
Injectate Sensor
Hoursing
Delivery System
In-line Injectate
Temperature probe
Figure 16-2 CO Sensor Connections
5.
You may repeat the measurement if necessary.
6.
When measurement(s) is finished, open WINDOW FOR C.O. EDIT to edit the
measured data. The rest of the sections include detailed procedure of editing.
NOTE
The blood temperature alarm is disabled during CO measurement. It will be
re-enabled when the measurement is over.
16-3