Shenzhen PM-9000 Express Manuel de l'opérateur Page 201

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 339
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 200
Temperature alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Feline 39.5 37.5
T1 and T2 Canine 39.5 38
Other animals 39.5 37.5
Feline 2 /
TD Canine 2 /
Other animals 2 /
14.4 Maintenance and Cleaning
WARNING
Before cleaning the monitor or the probe, make sure the equipment is turned off
and disconnected from AC power.
Reusable Temperature Probes
The temperature probe should not be heated to a temperature over 100. (212.). It is
only able to bear the temperature between 80 and 100. (176 to 212.) for a short time.
The probe must not be disinfected in steam.
Only detergents containing alcohol can be used for disinfection.
The rectal probes should be used, if possible, in conjunction with a protective rubber
cover.
To clean the probe, hold the tip with one hand and with the other hand rub the probe
down in the direction of the connector using a moist lint-free cloth.
NOTE
Disposable temperature probes must not be re-sterilized or reused.
Disposable temperature probes must be recycled or disposed of properly.
14-4
Vue de la page 200
1 2 ... 196 197 198 199 200 201 202 203 204 205 206 ... 338 339

Commentaires sur ces manuels

Pas de commentaire