Shenzhen PM-9000 Express Manuel de l'opérateur Page 262

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 339
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 261
19.3 Disinfection and Sterilization
Sterilization or disinfection may cause damage to the equipment. We recommend the
sterilization and disinfection are contained in the hospital’s servicing schedule only when
necessary. The equipment should be cleaned prior to sterilization and disinfection.
Recommended sterilization material: Alcohol based (Ethanol 70%, Isopropanol 70%), and
aldehyde based.
WARNING
Disinfection or sterilization may cause damage to the equipment; therefore,
when preparing to disinfect or sterilize the equipment, consult your hospital’s
infection controllers or professionals.
The cleaning solutions above can only be used for general cleaning. If you use
them to control infections, we shall assume no responsible for the effectiveness.
NOTE
ALWAYS dilute the solutions according to the manufacturer’s suggestions and
adopt lower concentration if possible.
NEVER submerge the equipment into water or any solution, or pour water or
any solution on the equipment;
ALWAYS wipe off all the excess liquids on the equipment surface and accessory
surface with a dry cloth;
Never use EtO and formaldehyde to disinfect.
Never permit high-pressure and high-temperature disinfection of the equipment
and accessories.
19-3
Vue de la page 261
1 2 ... 257 258 259 260 261 262 263 264 265 266 267 ... 338 339

Commentaires sur ces manuels

Pas de commentaire