6.3.5 Status Switchover
1.
In the Normal status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused status,
or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced status.
2.
In the Alarms Paused status,
Press the SILENCE key for less than 2 seconds to switch to the Normal status, or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced status.
3.
In the System Silenced status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused status,
or
Press the SILENCE key for 2 seconds or more to switch to the Normal status.
4.
In the Alarms Silenced status,
Enter the ALARM SETUP menu and set ALARM VOL to any setting other than 0.
The alarm sound is restored.
Enter the USER MAINTAIN menu and set ALARM SOUND OFF to DISABLE. The
system restores the default alarm volume.
6.4 Latching Alarms
As described in 4.7 Maintenance, the parameter alarm type can be set to either LATCH or
UNLATCH.
If the parameter alarm type is set to LATCH, before or during the occurence of a parameter
alarm, the alarm message will be latched even if the initial alarm condition has ceased. The
alarm messsage continues to be displayed, but the alarming modes change as follows:
The measured parameter value and the upper or lower parameter limit stop flashing.
The generated time of the alarm is displayed behind the alarm message in the
physiological alarms area.
If the parameter alarm type is set to UNLATCH, the monitor stops giving any indication for
this alarm when the initial alarm condition has ceased.
6-7